Acrylic Blood Tube Rack: Lab Bench Spec

A hospital laboratory specification for an acrylic blood tube rack usually runs to a single line: 50 place, 16 mm tubes, autoclavable. That last word is the one that decides the material, and it is the one most often written without checking whether the rack will ever see steam.

Where the answer is chemical disinfection, cast acrylic is the better bench material. Where it is steam, the rack has to be polypropylene and the acrylic version should not be quoted at all.

Hole diameter and the acrylic blood tube rack pattern

  • 13 mm tubes: 13.5 mm holes at 18 mm pitch
  • 16 mm tubes: 16.5 mm holes at 21 mm pitch
  • Well depth: 25 mm for 75 mm tubes, 40 mm for 100 mm tubes
  • Plate: 8 mm cast for 50 place, 10 mm cast above 100 place
  • Spill tray: 15 mm walls, 3 mm base, 5 mm below the plate
acrylic blood tube rack holding sample tubes on a clinical lab bench

Pitch matters as much as diameter. A rack at 18 mm pitch lets a gloved hand pick a single tube; at 15 mm the glove catches the neighbouring tube and the rack travels across the bench.

Cleaning, temperature and the acrylic blood tube rack life

Cast acrylic holds its shape to about 70 C continuously and distorts in a steam cycle at 121 C. Daily wipe down with 70 percent alcohol is harmless. Chlorinated disinfectants at bench strength will craze the surface over months, and a crazed rack shows hairline cracks around every hole, which is the point at which a laboratory replaces the whole set.

Spill containment and the acrylic blood tube rack policy

Many laboratory safety policies require a secondary container under open tubes. A 15 mm wall tray under the plate meets that requirement without raising the working height, and it doubles as the carrier when a full rack moves between benches.

River People cuts racks to your tube pattern with drilled and reamed holes, optional spill trays and colour coding by department. Send the tube diameter and count to oversales@rpplexiglass.com.

Send the tube diameter, tube count per rack and cleaning method to oversales@rpplexiglass.com for a quotation with one rack for bench trial.

Racks that leave the bench and travel to a ward are usually paired with PETG medical tray builds.

Frequently Asked Questions

Can an acrylic rack go in an autoclave?

No. Cast acrylic softens near 100 C and distorts above 80 C in a steam cycle. Use chemical disinfection, or specify polypropylene where steam sterilisation is mandatory.

What hole diameter suits a 13 mm tube?

13.5 mm. A 13 mm hole grips the tube and stops it dropping, while 14 mm lets a tube sit loose and lean against its neighbour.

How deep should the well be?

25 mm for a 75 mm tube, which holds the tube against a knock without hiding the label.